HPLC System 37-HPS100

HPLC System 37-HPS100

This HPLC system is equipped with high-pressure constant-current pump and UV detector. Pressure pulsation is controlled by diaphragmatic damper in seconds. The UV detectors helps in identification of samples and elutes that are colourless, odourless in nature. There is auto boot detection function available in this HPLC system.

High pressure pump Single pump
Flow rate range 0.001 - 9.999 mL / min (0.001 mL / min increment)
Pressure range ≤ 0.5 MPa (0 - 42 MPa)
Flow rate precision RSD < 0.06 %
Pulsation ≤ 0.1 MPa (flow rate 1 ml / min, Pressure 5 - 10 MPa)
Qualitative analysis RSD 6 ≤ 0.1 % (naphthalene methyl alcohol solvent)
Quantitative repeatability RSD 6 ≤ 0.3 % (naphthalene methyl alcohol solvent)
Flow rate setting error Ss ≤ 2 % (1 mL / min water, 5 - 10 MPa, room temperature)
Mixing accuracy ± 1 %
Mixing precision ± 0.2 %
Size (W × H × D) 450 × 300 × 160 mm
UV Detector a:9:{s:16:"Wavelength range";s:12:"190 - 680 nm";s:9:"Flow cell";s:5:"8 µL";s:19:"Spectrum band width";s:4:"8 nm";s:14:"Baseline noise";s:46:"≤ ± 0.25 × 10-5 AU (empty cell)";s:19:"Wavelength accuracy";s:7:"± 1 nm";s:14:"Baseline drift";s:45:"≤ ± 0.4 × 10-4 AU (empty cell)";s:20:"Wavelength precision";s:8:"< 0.1 nm";s:18:"Limit of detection";s:64:"3 × 10-9 g / mL (naphthalene methyl alcohol solvent)";s:18:"Size (W × H × D)";s:20:"450 × 300 × 160 mm";}
  • Limit baseline noise can be ensured using electronic pulse suppression technique
    High pressure single pump
    Easy to change gasket ring & automatic washing function for the column
    User-friendly software for smooth operations
  • The HPLC has applications in separation of mixture of compounds in chemical analysis, poison analysis, environmental analysis, protein food detection, medicine testing, and polymer analysis.

Analytical HPLC

Analytical HPLC 37-AHP100

  • Pump Specifications :
    Model No HPP-A1
    Flow rate range 0.0001 ~ 10 ml / min
    Maximum pressure 50 MPa , 0 ~ 2 ml / min
    Column size 4.6 x 250 mm11
    Sample size 5 µm
    Pressure pulsation ≤ 1 % at 10 MPa
    Flow rate accuracy ± 0.15 % at 1 ml / min
    Flow rate repeatability % ≤ 0.1
    Gradient accuracy % ± 1.5
    Gradient repeatability % ≤ 0.2
    Dimension 460 x 310 x 180 mm
    Power supply 90 V
    Weight 11.6 kg
  • Detector Specifications :
    Model No HPD-A20 HPD-A21
    Wavelength range Single wavelength ( 190 to 800 nm ) Dual wavelength ( 190 to 800 nm )
    Wavelength accuracy ± 0.5 nm ± 2 nm
    Wavelength repeatability 0.15 nm 0.3 nm
    Band width 8 nm 8 nm
    Optical path 10 nm 10 nm
    Deuterium lamp 190 ~ 400 nm 190 ~ 400 nm
    Tungsten lamp 400 ~ 800 nm 400 ~ 800 nm
    Baseline noise ( Static ) Au ≤ ± 0.35 x 10 -5 ≤ ± 0.5 x 10 -5
    Baseline drift ( Static ) Au / h ≤ 1 x 10 -4 ≤ 1.5 x 10 -4
    Minimum detection limit ≤ 4 x 10 -8 ≤ 4 x 10 -8
    Power 100 W 100 W
    Power supply 90 V / 50 ~ 60 Hz 90 V / 50 ~ 60 Hz
    Dimension 460 x 315 x 180 mm 460 x 315 x 180 mm
    Weight 9.5 kg 9.5 kg

Analytical HPLC 37-AHP101

  • Pump Specifications :
    Model No HPP-B1
    Flow rate range 0.0001 ~ 10 ml / min
    Maximum pressure 42 MPa , 0 ~ 5 ml / min
    Column size 4.6 x 250 mm11
    Sample size 5 µm
    Pressure pulsation ≤ 1.5 % at 10 MPa
    Flow rate accuracy ± 0.2 % at 1 ml / min
    Flow rate repeatability % ≤ 0.15
    Gradient accuracy % ± 2
    Gradient repeatability % ≤ 0.2
    Dimension 450 x 260 x 190 mm
    Power supply 90 V ~ 240 V
    Weight 11.2 kg
  • Detector Specifications :
    Model No HPD-B2
    Wavelength range 190 to 800 nm
    Wavelength accuracy ± 0.75 nm
    Wavelength repeatability 0.2nm
    Band width 8 nm
    Optical path 10 nm
    Deuterium lamp 190 ~ 400 nm
    Tungsten lamp 400 ~ 800 nm
    Baseline noise ( Static ) Au ≤ ± 0.5 x 10 -5
    Baseline drift ( Static ) Au / h ≤ 1.5 x 10 -4
    Minimum detection limit ≤ 4 x 10 -8
    Power 100 W
    Power supply 90 V / 50 ~ 60 Hz
    Dimension 460 x 260 x 190 mm
    Weight 9.2 kg

Preparative HPLC

Preparative HPLC 37-PHP100

  • Dimension : 750 x 860 x 1280 mm ( 100 ml , 250 ml , 500 ml , 1000 ml ) , 810 x 950 x 1310 mm ( 3 L )
  • Wetted material : 316 L and PTFE

Semi-prep HPLC

Semi-prep HPLC 37-SPH100

  • Pump Specifications :
    Model No SPH-A10 SPH-A11 SPH-A12 SPH-A13
    Flow rate range 0.001 ~ 50 ml / min 0.001 ~ 50 ml / min 0.001 ~ 100 ml / min 0.001 ~ 100 ml / min
    Maximum pressure 30 MPa 30 MPa 15 MPa 10 MPa
    Flow rate accuracy ± 0.35 ± 0.5 ± 0.5 ± 0.5
    Flow rate repeatability % ≤ 0.1 ≤ 0.15 ≤ 0.1 ≤ 0.15
    Column size 10 x 250 mm 10 x 250 mm 10 x 250 mm 10 x 250 mm
    Sample loading 35 ~ 60 mg 35 ~ 60 mg 35 ~ 60 mg 35 ~ 60 mg
    Pressure pulsation 1 % at 10 MPa 1.5 % at 10 MPa 1.5 % at 10 MPa 1.5 % at 10 MPa
    Flow rate repeatability % ≤ 0.1 ≤ 0.15
    Gradient accuracy % ± 2 ± 2 ± 2 ± 2
    Gradient repeatability % ≤ 0.2 ≤ 0.2 ≤ 0.2 ≤ 0.2
    Dimension 465 x 315 x 180 mm 455 x 265 x 190 mm 465 x 315 x 180 mm 465 x 315 x 180 mm
    Power 150 W 150 W 150 W 150 W
    Power supply 90 V ~ 240 V/ 50 Hz 90 V ~ 240 V/ 60 Hz 90 V ~ 240 V/ 50 Hz 90 V ~ 240 V/ 60 Hz
    Weight 11.6 kg 11.2 kg 11.6 kg 11.6 kg
  • Detector Specifications :
    Model No SPH-A20 SPH-A21 SPH-A22
    Wavelength range 190 to 800 nm 190 to 800 nm 190 to 800 nm
    Wavelength accuracy ± 0.5 nm ± 1 nm ± 0.75 nm
    Wavelength repeatability 0.15 nm 0.3 nm 0.2 nm
    Band width 8 nm 8 nm 8 nm
    Optical path Semi flow / analytical Semi flow / analytical Semi flow / analytical
    Baseline noise ( Static ) Au ≤ ± 0.35 x 10 -5 ≤ ± 0.5 x 10 -5 ≤ ± 0.5 x 10 -5
    Baseline drift ( Static ) Au / h ≤ 1 x 10 -4 ≤ 1.5 x 10 -4 ≤ 1.5 x 10 -4
    Minimum detection limit ≤ 4 x 10 -8 ≤ 4 x 10 -8 ≤ 4 x 10 -8
    Power 150 W 150 W 150 W
    Power supply 90 V / 50 ~ 60 Hz 90 V / 50 ~ 60 Hz 90 V / 50 ~ 60 Hz
    Dimension 465 x 315 x 180 mm 465 x 315 x 180 mm 460 x 265 x 190 mm
    Weight 9.5 kg 9.5 kg 9.2 kg
Enquiry Form