Plasma Autoclave 26-PAC201

Plasma Autoclave 26-PAC201

Plasma autoclave comes with sliding door, rectangular chamber design, microprocessor PLC controlled system and 130 L volume capacity. Adopted with excellent thermal conductivity ensures 100% maintenance of gaseous hydrogen peroxide. Equipped with high and low temperature alarm system, pressure and temperature sensors, H2O2 resistant rotary vane vacuum pump for quality performance. The maximum storage space is obtained by loading items on the two layers of aluminum alloy punching plate.

$18720
Chamber Volume 130 L
Sterilization temperature 50 ± 5°C
Working pressure -50 Pa
Sterilizing agent 60 % (H2O2) hydrogen peroxide
Sterilizing time 45 to 50 min
Installation environment temperature 5 to 45°C
Vacuum speed Absorb to – 50 Pa in five minutes
High anti-pressure valve 10 Pa
Door type Rectangular type, sliding door
Operation height (for chamber design) ≤ 160 cm
Chamber material Anti-corrosion type SUS 304 stainless steel
Chamber temperature heating power 2.75 kW, pre-heating time 30 mins
Electrode material Anti-corrosion type SUS 304 stainless steel
Sealing door material Anti-corrosion type SUS 304 stainless steel
Carton material Carbon steel baking plastic powder and ABS
Storage U disk capacity ≥ 8 G
Air filter accuracy ≤ 0.2 µm
Chamber dimension (L*W*H) 750 × 450 × 400 mm
Packaging dimension (L*W*H) 1070 × 810 × 1780 mm
Power 4.35 kW
Power supply heating method AC220 V, 50 Hz
Gross weight 360 kg
  • 7 inch touch screen LCD display
    Electric lift door open mode
    Fully automatic control system
    Large storage capacity
    High strength rotary vane vacuum pump
    Exhaust oil mist filtration system
    Door protection switch function
    Adapted with anti-explosion and fire resistance heating system
    Micro printer (data reviewing) attachment
    Manual setting for different sterilization programs
    Use friendly appearance
  • Accessories No. Name
    1 Hydrogen Peroxide (6-10 g/ bottle)
    2 Indicative tape (20 × 25 mm/ roll)
    3 Biological indicator (25 pieces/ box)
    4 Chemical indicative card (200 pieces/ box)
    5 Chemical indicative label
    6 Sterilization packing bag
    7 Non-woven fabrics
  • Used in the hospital, pharmaceutical factory, lab, medical institutions for low temperature sterilization of heat sensitive materials.

Customer Reviews

Horizontal Autoclave

Medical Autoclave

Portable Autoclave

Tabletop Autoclave

Vertical Autoclave

Tabletop Steam Autoclave

Tabletop Steam Autoclave TSA100

  • Capacity : 80 L
  • Sterilization Temperature : 134 ℃
$5380

Waste Autoclave

Waste Aautoclave 26-WAC100

  • Capacity : 1200 L
  • Temperature Range : 115 °C to 139 °C
$26860

Enquiry Form

Captcha Image